Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Effect of Exposure Parameters on Metal Artifacts in Cone Beam Computed Tomography
Mehrdad Panjnoush1, Yasaman Kheirandish1, Pooneh Mohseni Kashani2
1Assistant Professor, Department of Oral and Maxillofacial Radiology, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran.
Objectives:
The aim of this study was to assess the effect of tube current, kilovoltage peak (kVp), metal type, and the position of metal objects on metal artifacts in cone-beam computed tomography (CBCT) images.
Materials And Methods:
Titanium and cobalt-chromium rods were fabricated and placed in a dry human mandible. Samples were scanned using a Promax 3D CBCT unit with different milli-amperages and kVp. The artifacts induced by metal objects were evaluated using the Image J software in four regions of interest (ROIs) on each image.
Results:
A higher kVp decreased artifacts of the buccal surface of the rods in 97% of the cases (P=0.046) but did not affect the severity of artifacts between the two metal objects (P>0.05). Increasing the tube current had no effect on metal artifacts in 93% of the cases (P>0.05). Artifacts induced by a cobalt-chromium alloy were more severe than those with titanium (P<0.001). Artifacts were more intense in the buccal surface of anterior rods compared to the posterior rods (P<0.003).
Conclusions:
Tube voltage, metal type and the position of metal objects affected the severity of metal artifacts on CBCT images. The metal type had the greatest effect on metal artifact intensity in this study.
Related Concept Videos
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...