Monitoring Fetal Heart Rate during Labor: A Comparison of Three Methods

Tammy Y Euliano1, Shalom Darmanjian2, Minh Tam Nguyen2

  • 1Department of Anesthesiology, University of Florida College of Medicine, Gainesville, FL, USA; Department of Obstetrics and Gynecology, University of Florida College of Medicine, Gainesville, FL, USA.

Journal of Pregnancy
|April 11, 2017
PubMed

The purpose of the study was to compare the accuracy of a noninvasive fetal heart rate monitor with that of ultrasound, using a fetal scalp electrode as the gold standard, in laboring women of varying body habitus, throughout labor and delivery. Laboring women requiring fetal scalp electrode were monitored simultaneously with the investigational device (noninvasive fetal ECG), ultrasound, and fetal scalp electrode. An algorithm extracted the fetal heart rate from the noninvasive fetal ECG signal. Each noninvasive device recording was compared with fetal scalp electrode with regard to reliability by positive percent agreement and accuracy by root mean squared error. Seventy-one women were included in this analysis. Positive percent agreement was 83.4 ± 15.4% for noninvasive fetal ECG and 62.4 ± 26.7% for ultrasound. The root mean squared error compared with fetal scalp electrode-derived fetal heart rate was 4.8 ± 2.0 bpm for noninvasive fetal ECG and 14.3 ± 8.2 bpm for ultrasound. The superiority of noninvasive fetal ECG was maintained for stages 1 and 2 of labor and increases in body mass index. Compared with fetal scalp electrode-derived fetal heart rate, noninvasive fetal ECG is more accurate and reliable than ultrasound for intrapartum monitoring for stages 1 and 2 of labor and is less affected by increasing maternal body mass index. This confirms the results of other workers in this field.

Related Concept Videos

Temperature Measurement Sites01:14

Temperature Measurement Sites

A thermometer measures body temperature. The common sites for measuring body temperature are the oral cavity, axillary region, temporal artery, and skin surface, such as the forehead, abdomen, and axilla. True core body temperature is assessed in the rectum, tympanic membrane, pulmonary artery, esophagus, and urinary bladder.
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Pulse rhythm01:30

Pulse rhythm

Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)01:15

Assessment of blood pressure in brachial artery(one-step method)

This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
Fetal Circulation01:14

Fetal Circulation

Fetal circulation is a unique system that facilitates the exchange of gases, nutrients, and waste products between the developing fetus and the mother. This intricate process takes place through a special organ called the placenta.
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...