Computed tomography pulmonary angiography using a 20% reduction in contrast medium dose delivered in a multiphasic injection
Related Experiment Video
Updated: Sep 30, 2026

Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
Mitchell Chen1, Gaith Mattar1, Jamal A Abdulkarim1
1Mitchell Chen, Jamal A Abdulkarim, Department of Radiology, George Eliot Hospital NHS Trust, Nuneaton, Warwickshire CV10 7DJ, United Kingdom.
Aim:
To evaluate the feasibility of reducing the dose of iodinated contrast agent in computed tomography pulmonary angiography (CTPA).
Methods:
One hundred and twenty-seven patients clinically suspected of having pulmonary embolism underwent spiral CTPA, out of whom fifty-seven received 75 mL and the remaining seventy a lower dose of 60 mL of contrast agent. Both doses were administered in a multiphasic injection. A minimum opacification threshold of 250 Hounsfield units (HU) in the main pulmonary artery is used for assessing the technical adequacy of the scans.
Results:
Mean opacification was found to be positively correlated to patient age (Pearson's correlation 0.4255, P < 0.0001) and independent of gender (male:female, 425.6 vs 450.4, P = 0.34). When age is accounted for, the study and control groups did not differ significantly in their mean opacification in the main (436.8 vs 437.9, P = 0.48), left (416.6 vs 419.8, P = 0.45) or the right pulmonary arteries (417.3 vs 423.5, P = 0.40). The number of sub-optimally opacified scans (the mean opacification in the main pulmonary artery < 250 HU) did not differ significantly between the study and control groups (7 vs 10).
Conclusion:
A lower dose of iodine contrast at 60 mL can be feasibly used in CTPA without resulting in a higher number of sub-optimally opacified scans.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography