The tip-apex distance in the X-Bolt dynamic plating system.

M A Fernandez1, A Aquilina2, J Achten2

  • 1University of Oxford, NDORMS, John Radcliffe Hospital, Oxford OX3 9DU and Specialty Registrar, Trauma and Orthopaedic Surgery, University Hospital Coventry, Clifford Bridge Road, Coventry, CV2 2DX, UK miguel.fernandez@ndorms.ox.ac.uk.

Bone & Joint Research
|April 14, 2017
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Co-designing roadmaps for promoting culturally relevant food in long-term care homes for the African, Caribbean and Black populations in Canada: a multi-method, participatory action research study protocol.

BMC nutrition·2026
Same author

Wear influences mercury concentrations and stable isotope values in penguin feathers.

Environmental toxicology and chemistry·2025
Same author

Facilitating the transition from hospital to home after hip fracture surgery: a qualitative study from the HIP HELPER trial.

BMC geriatrics·2024
Same author

Correction: Complex organisational factors influence multidisciplinary care for patients with hip fractures: a qualitative study of barriers and facilitators to service delivery.

BMC musculoskeletal disorders·2024
Same author

Impact of colonoscopy on health-related quality of life: findings from the RECEDE study.

Health and quality of life outcomes·2024
Same author

Analytical corrections for B<sub>1</sub>-inhomogeneity and signal decay in multi-slice 2D spiral hyperpolarized <sup>129</sup>Xe MRI using keyhole reconstruction.

Magnetic resonance in medicine·2024

The X-Bolt dynamic hip plating system showed a greater tip-apex distance compared to the sliding hip screw. However, this did not lead to an increased risk of fixation failure or lag screw cut-out in unstable hip fractures.

Area of Science:

  • Orthopaedic surgery
  • Trauma surgery
  • Biomedical engineering

Background:

  • Trochanteric hip fractures are common, with fixation failure a significant complication.
  • Lag screw cut-out is a mode of failure, particularly in unstable fractures and with increased tip-apex distance (TAD).
  • The X-Bolt dynamic hip plating system aims to reduce cut-out risk, but concerns exist regarding its TAD.

Purpose of the Study:

  • To compare the tip-apex distance (TAD) between the X-Bolt system and the sliding hip screw (SHS).
  • To evaluate the incidence of lag screw cut-out with both devices in unstable trochanteric hip fractures.

Main Methods:

  • A randomized controlled trial (WHiTE One trial) involving 93 patients with unstable trochanteric hip fractures.
  • Measurement of TAD from intra-operative radiographs for both SHS and X-Bolt groups.
Keywords:
Hip fractureTip-apex distanceX-Bolt

Related Experiment Videos

  • Recording of cut-out incidence at a median follow-up of 17 months.
  • Main Results:

    • The X-Bolt group had a significantly increased mean TAD (20.8 mm) compared to the SHS group (17.1 mm) (mean difference 3.7 mm, p = 0.001).
    • Despite the increased TAD, no cut-outs were observed in the X-Bolt group (0%) compared to three (6%) in the SHS group.
    • The difference in cut-out incidence was not statistically significant at the 17-month follow-up.

    Conclusions:

    • The X-Bolt dynamic hip plating system is a safe implant for treating unstable trochanteric hip fractures.
    • Concerns regarding achieving a minimal tip-apex distance with the X-Bolt may be unwarranted, as it does not appear to increase the risk of cut-out.
    • The X-Bolt system demonstrates comparable or improved outcomes regarding fixation failure compared to the sliding hip screw.