Evaluation of Different Estimation Methods for Accuracy and Precision in Biological Assay Validation
1MEDIMMUNE LLC, GAITHERSBURG, MD yub@medimmune.com.
PDA Journal of Pharmaceutical Science and Technology
|April 19, 2017
Summary
Biological assays (bioassays) are crucial for drug development and quality control. This study clarifies bias and precision definitions and proposes alternative relative standard deviation (RSD) calculation methods, outperforming USP standards for accuracy.
Area of Science:
- Biostatistics
- Pharmacometrics
- Analytical Chemistry
Background:
- Biological assays (bioassays) are vital for estimating substance potency and monitoring biopharmaceutical products.
- Key performance metrics include relative accuracy (bias) and precision, with existing guidelines (USP <1033>) offering unclear definitions and methods.
- The utility of data transformations, like log-transformation, for intermediate precision estimation remains ambiguous, complicating method selection for analysts.
Purpose of the Study:
- To provide rigorous definitions for bias and precision in bioassay analysis.
- To introduce and evaluate three alternative methods for calculating relative standard deviation (RSD).
- To compare the performance of proposed RSD methods against the USP <1033> standard, particularly concerning bias and root-mean-square error (RMSE).
Main Methods:
- Development of precise definitions for bioassay bias and precision.
- Implementation and comparison of three novel RSD calculation methods.
- Statistical analysis to evaluate bias and RMSE across different methods and data distributions.
Main Results:
- All evaluated methods performed similarly when relative standard deviation (RSD) was ≤10%.
- The USP <1033> method exhibited larger bias and RMSE compared to the proposed methods for high-variation data.
- Estimates based on the original data scale performed well for moderately skewed data; log-transformation is suitable for noticeably skewed data.
Conclusions:
- The USP <1033> method for RSD calculation is not recommended for routine bioassay analysis due to potential bias and RMSE issues.
- The original scale method is preferred for estimating precision in bioassays with moderate skewness.
- Log-transformation offers a viable alternative for precision estimation with noticeably skewed bioassay data.