Evaluation of the truebeam machine performance check (MPC): mechanical and collimation checks
Michael P Barnes1,2,3, Peter B Greer1,3
1Department of Radiation Oncology, Calvary Mater Hospital Newcastle, Waratah, NSW, Australia.
Journal of Applied Clinical Medical Physics
|April 19, 2017
Summary
Machine Performance Check (MPC) is a suitable tool for daily quality assurance (QA) of TrueBeam linear accelerators. It accurately verifies geometric and beam performance, showing good agreement with established QA methods.
Area of Science:
- Medical Physics
- Radiation Oncology
- Quality Assurance
Background:
- TrueBeam linear accelerators require regular performance verification.
- Traditional quality assurance (QA) methods can be time-consuming.
- Automated tools like Machine Performance Check (MPC) offer potential for efficient QA.
Purpose of the Study:
- To evaluate the performance of MPC's geometric tests for TrueBeam linacs.
- To assess MPC's accuracy in verifying beam collimation (MLC, jaws) and mechanical systems (gantry, collimator).
- To compare MPC's results with routine departmental QA tests.
Main Methods:
- Compared MPC's Multi-Leaf Collimator (MLC) tests against Picket Fence analysis.
- Compared MPC's jaw position checks against EPID-based methods, light field/graph paper, and Daily QA3.
- Compared MPC's gantry and collimator checks against spirit level and Picket Fence analysis.
- Assessed MPC's sensitivity by introducing deliberate calibration errors.
Main Results:
- MPC MLC tests agreed with Picket Fence analysis within 0.3 mm.
- MPC jaw checks agreed with in-house EPID measurements within 0.2 mm.
- All MPC parameters demonstrated accurate sensitivity to introduced calibration errors.
- MPC results showed good agreement with established routine QA procedures.
Conclusions:
- Machine Performance Check (MPC) is a suitable automated tool for daily QA of TrueBeam linacs.
- MPC accurately verifies geometric and beam collimation performance.
- MPC demonstrates reliability and sensitivity for quality assurance in radiation oncology.