Related Experiment Video
Updated: Mar 3, 2026

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Multicenter Investigation of the Effective Dose for Whole-body CT Scanning
Toshimitsu Sato1, Yukihiro Saito1, Nobutoshi Saito1
1Department of Radiology, Yamagata University Hospital.
Objective:
To study whether the actual radiation exposure is different between computed tomography (CT) scanners and medical centers when the same patient is scanned, we investigated the actual effective doses for a whole body (Chest-Pelvis) CT scan in a multicenter study.
Materials And Methods:
Data from subjects were collected using 12 CT scanners at six medical centers in Yamagata city. Effective-dose data were acquired by scanning the same phantom (ATOM Dosimetry Phantoms Model702-B) using 120 kV tube voltage. Effective doses were calculated using corrected data from a radiophotoluminescent glass dosimeter (GD-302M). GD-302M had energy- dependent issues, which needed to be corrected. Also, differences in sensitivity based on arrangement within the phantom were insignificant.
Results:
The mean effective energy was 48.6 keV (range, 42.5-55.4 keV), and the mean effective dose was 16.3 mSv (range, 8.9-26.0 mSv). The mean effective dose with a hybrid type iterative reconstruction was 10.7 mSv (range, 8.9-16.4 mSv), but the mean effective dose without any iterative reconstruction was 20.3 mSv (range, 16.2-26.0 mSv). We found an approximate linear correlation between dose length product (DLP) on operation consoles and the effective dose.
Conclusion:
We suggest that the actual radiation exposure was different at each medical center when the same patient is scanned.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT