Comparative Study of Two ELISA Kits for Estimation of Antibodies to Hepatitis B Virus in Human Normal Immunoglobulin and Specific Immunoglobulin Intended for Intravenous or Intramuscular Use
Related Experiment Video
Updated: Mar 3, 2026

Author Spotlight: Advancements and Challenges in Hepatitis B Virus Detection
Published on: December 15, 2023
J P Prasad1, Y Madhu1, Arushi Mandhan1
1Blood Products Laboratory, National Institute of Biologicals, A-32, Sector 62, Noida, 201 309 Uttar Pradesh India.
Abstract:
Current study is conducted to evaluate method verification of two locally available kits manufactured by DSI & BIORAD for quantitative estimation of Hepatitis B virus antibodies in human normal immunoglobulin by using International standard of National Institute of Biological Standards and Control. Four analyst perform five sets of test in duplicate analysing accuracy, precision, and limit of detection, sensitivity and specificity. Our results suggest that both DSI and BIORAD kits fulfil the validation criteria and are sensitive to detect up to 10 mIU concentration precisely and accurately. DSI kit is more precise at concentration 100 mIU and economically 4-5 times cheaper in local market; on the other hand, BIORAD kits provide larger detection range up to 1000 mIU.