Direct measurement of clinical mammographic x-ray spectra using a CdTe spectrometer
Josilene C Santos1, Alessandra Tomal2, Tânia A Furquim1
1Instituto de Física, Universidade de São Paulo, São Paulo, SP, 05508-090, Brazil.
Purpose:
To introduce and evaluate a method developed for the direct measurement of mammographic x-ray spectra using a CdTe spectrometer. The assembly of a positioning system and the design of a simple and customized alignment device for this application is described.
Methods:
A positioning system was developed to easily and accurately locate the CdTe detector in the x-ray beam. Additionally, an alignment device to line up the detector with the central axis of the radiation beam was designed. Direct x-ray spectra measurements were performed in two different clinical mammography units and the measured x-ray spectra were compared with computer-generated spectra. In addition, the spectrometer misalignment effect was evaluated by comparing the measured spectra when this device is aligned relatively to when it is misaligned.
Results:
The positioning and alignment of the spectrometer have allowed the measurements of direct mammographic x-ray spectra in agreement with computer-generated spectra. The most accurate x-ray spectral shape, related with the minimal HVL value, and high photon fluence for measured spectra was found with the spectrometer aligned according to the proposed method. The HVL values derived from both simulated and measured x-ray spectra differ at most 1.3 and 4.5% for two mammography devices evaluated in this study.
Conclusion:
The experimental method developed in this work allows simple positioning and alignment of a spectrometer for x-ray spectra measurements given the geometrical constraints and maintenance of the original configurations of mammography machines.
More Related Videos
11:42Measurement of Total Calcium in Neurons by Electron Probe X-ray Microanalysis
Published on: November 20, 2013
10:42In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography CT and Light Microscopy LM Correlated with Scanning Electron Microscopy SEM
Published on: June 16, 2016
Related Concept Videos
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....