Lower Esophageal Sphincter Augmentation for Gastroesophageal Reflux Disease: The Safety of a Modern Implant
C Daniel Smith1, Robert A Ganz2, John C Lipham3
11 Esophageal Institute of Atlanta , Atlanta, Georgia .
Introduction:
Use of the magnetic sphincter augmentation device (MSAD) for gastroesophageal reflux disease (GERD) is increasing. As this innovative treatment for GERD gains widespread use and adoption, an assessment of its safety since U.S. market introduction is presented.
Methods:
Events were collected from the Manufacturer and User Facility Device Experience (MAUDE) database, which reports events submitted to the Food and Drug Administration (FDA) of suspected device-associated deaths, serious injuries, and malfunctions. The reporting period was from March 22, 2012 (FDA approval) through May 31, 2016, and included only events occurring in the United States. Additional information was provided by the manufacturer, allowing calculation of implant rates and durations.
Results:
An estimated 3283 patients underwent magnetic sphincter augmentation (165 surgeons at 191 institutions). The median implant duration was 1.4 years, with 1016 patients implanted for at least 2 years. No deaths, life-threatening events, or device malfunctions were reported. The overall rate of device removal was 2.7% (89/3283). The most common reasons for device removal were dysphagia (52/89) and persistent reflux symptoms (19/89). Removal for erosion and migration was 0.15% (5/3283) and 0% (0/3283), respectively. There were no perforations. Of the device removals, 57.3% (51/89) occurred <1 year after implant, 30.3% (27/89) between 1 and 2 years, and 12.4% (11/89) >2 years after implant. The rate of device removal and erosion with an implant duration >2 years were 1.1% (11/1016) and 0.1% (1/1016), respectively. All device removals and erosions were managed nonemergently, with no complications or long-term consequences.
Conclusions:
During a 4-year period in more than 3000 patients, no unanticipated MSAD complications have emerged, and there is no data to suggest a trend of increased events over time. The presentation and management of device-related issues have been less complicated than revisions for laparoscopic fundoplication or other interventions for GERD. MSAD is considered safe for the widespread treatment of GERD.
More Related Videos
08:25Construction of a Wireless-Enabled Endoscopically Implantable Sensor for pH Monitoring with Zero-Bias Schottky Diode-based Receiver
Published on: August 27, 2021
04:05Endoscopic Vacuum Therapy for the Treatment of Anastomotic Leakage after Total Gastrectomy with Esophagojejunostomy
Published on: August 22, 2025
Related Concept Videos
Gastroesophageal Reflux Disease II: Clinical Features and Management
Clinical Manifestations
GERD presents itself in a multitude of ways, with symptoms varying from person to person. The hallmark symptoms are...
Barrett Esophagus-I: Introduction
This constant acid exposure transforms the esophagus's pink mucosal lining (stratified squamous epithelium) into a type of lining more similar...
Barrett Esophagus-II: Clinical Manifestations and Management
To diagnose Barrett's esophagus, healthcare providers often recommend an endoscopy for those showing symptoms of acid reflux. The procedure entails...
Esophageal Strictures-I: Introduction
Etiology
The primary cause of esophageal strictures is long-standing gastroesophageal reflux disease (GERD), accounting for about 70 to 80% of adult cases. Chronic acid reflux can lead to injury and scarring of the esophageal lining, culminating in...
Esophageal Strictures-II: Clinical Features and Management
Healthcare providers should gather a comprehensive medical history and conduct a physical examination for diagnosis. If esophageal stricture is...
Esophageal Achalasia