Occurrence of simultaneous cathodal-anodal capture with left ventricular quadripolar leads for cardiac
Eraldo Occhetta1, Gabriele Dell'Era1, Ailia Giubertoni1
1Department of Cardiology, University Hospital 'Maggiore della Carità', Corso Mazzini 18, 28100 Novara, Italy.
Aims:
The occurrence of left ventricular (LV) anodal activation during pacing with modern multipolar cardiac resynchronization therapy (CRT) systems has never been reported. The aim of our study was to demonstrate, by means of electrocardiogram (ECG) analysis, the occurrence of simultaneous cathodal-anodal LV capture with quadripolar LV leads.
Methods And Results:
We studied 10 first-time recipients of a CRT device equipped with a quadripolar LV lead. During follow-up, standard supine 12-lead ECGs were obtained in available cathode-to-anode LV pacing configurations with a pulse amplitude equal to twice the pacing threshold. The occurrence of simultaneous cathodal-anodal LV capture was defined as the presence of variations in electrocardiographic ventricular activation (EVA) when the distal tip (cathode)-to-device can (anode) pacing configuration was compared with the distal tip (cathode)-to-proximal ring (anode) configuration. In eight patients, we found differences in EVA when different LV sites were paced through the unipolar LV tip and unipolar LV ring configurations. In these patients, a difference in EVA was detected in 61.5% (59 of 96) of the ECG leads (marked difference in 31.3%, slight difference in 30.2%). Changes in EVA between unipolar tip-to-can and bipolar tip-to-ring pacing that were suggestive of cathodal-anodal LV capture were found in six patients. In these patients, a total of 30 (41.7%) ECG leads showed a difference in EVA (marked difference in 20.8%, slight difference in 20.8%).
Conclusion:
In our experience, additional anodal capture by the proximal LV ring during LV pacing is provable in most recipients of a resynchronization device equipped with a multipolar LV lead.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...