Quantitative evaluation of metal artifacts using different CBCT devices, high-density materials and field of views
Marina Codari1,2, Karla de Faria Vasconcelos1,3, Laura Ferreira Pinheiro Nicolielo1
1OIC, OMFS IMPATH Research Group, Department Imaging & Pathology, Faculty of Medicine, KU Leuven and Oral and Maxillofacial Surgery, University Hospitals Leuven, Leuven, Belgium.
Objective:
To objectively compare the influence of different cone-beam computed tomography (CBCT) devices, high-density materials and field of views (FOVs) on metal artifact expression.
Material And Methods:
For this in vitro study, three customized acrylic resin phantoms containing high-density materials cylinders: titanium, copper-aluminum alloy and amalgam were scanned on three CBCT devices using high-resolution protocols, same voxel size (0.2 mm) and different FOVs. After fully automatic segmentation and image registration, the same region of interest was defined for the small and medium FOVs. The difference between the segmented and the real volume of the metal cylinders was assessed. Moreover for each segmented slice, the area difference between the segmented and the real axial section was determined. The artifacts on the background were measured as normalizing standard deviation of voxel values in the vicinity of the cylinder, in three different distances.
Results:
Considerable differences were observed in volume measurements for all CBCTs devices and materials for both FOV sizes (up to 67%). The slice per slice area analysis indicated higher artifacts at the edges of the metal cylinder. Within the materials, amalgam and titanium had, respectively, the worst and best artifact expression in all the CBCT devices. Standard deviation values varied differently between the three distances in each device.
Conclusion:
Our in vitro study showed that different CBCT devices, high-density materials and FOV should be considered while evaluating CBCT images. More carefully, diagnosis conclusions should be drawn in images containing amalgam and copper-aluminum alloy.
More Related Videos
10:10Three-Dimensional Particle Shape Analysis Using X-ray Computed Tomography: Experimental Procedure and Analysis Algorithms for Metal Powders
Published on: December 4, 2020
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...