Related Experiment Video
Updated: Mar 3, 2026

Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
In situ Hearing Tests for the Purpose of a Self-Fit Hearing Aid
Monique Boymans1, Wouter A Dreschler
1Academic Medical Center, Amsterdam, The Netherlands.
Abstract:
This study investigated the potential and limitations of a self-fit hearing aid. This can be used in the "developing" world or in countries with large distances between the hearing-impaired subjects and the professional. It contains an on-board tone generator for in situ user-controlled, automated audiometry, and other tests for hearing aid fitting. Twenty subjects with mild hearing losses were involved. In situ audiometry showed a test-retest reliability (SD <3.7 dB) that compared well with the precision of diagnostic audiometry using headphones. There was good correspondence (SD <5.2 dB) with traditional pure-tone audiometry. In situ loudness scaling yielded important information about suprathreshold perception, which will have an added value for the selection of compression and the selection of maximum power output to be allowed in hearing aids.
More Related Videos
04:32Sound Source Localization Testing in Single-sided Deafness Following Bone Conduction Intervention
Published on: December 20, 2024
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024