Effective dose estimation for oncological and neurological PET/CT procedures
Josep M Martí-Climent1,2, Elena Prieto3,4, Verónica Morán3
1Nuclear Medicine Department, Clínica Universidad de Navarra, 36, Pío XII Avenue, 31008, Pamplona, Spain. jmmartic@unav.es.
EJNMMI Research
|April 26, 2017
Summary
This study evaluated patient effective dose (ED) in PET/CT scans using various radiopharmaceuticals. CT significantly contributes to the dose in body scans, while the radiopharmaceutical is the main dose contributor in brain scans.
Area of Science:
- Nuclear Medicine
- Radiological Sciences
- Medical Imaging
Background:
- Patient effective dose (ED) varies across different Positron Emission Tomography/Computed Tomography (PET/CT) procedures.
- Understanding dose contributions from both PET radiopharmaceuticals and CT is crucial for radiation safety.
Purpose of the Study:
- To retrospectively evaluate patient effective dose (ED) for diverse PET/CT protocols using various radiopharmaceuticals.
- To determine the relative contributions of PET radiotracers and CT to the total ED in different scan types.
Main Methods:
- Retrospective review of 210 patient PET/CT studies across Torso, Whole Body (WB), Head and Neck Tumor (HNT), and Brain protocols.
- Utilized various radiopharmaceuticals including 18F-FDG, 11C-CHOL, 18F-FDOPA, 11C-MET, and 18F-florbetapir.
- Calculated ED using conversion factors applied to radiotracer activity and CT dose-length product.
Main Results:
- Mean ED ranged from 4.6 mSv (Brain PET/CT with 18F-FDOPA) to 20.0 mSv (WB-18F-FDG).
- CT was the primary dose contributor for body protocols (62-77% ED), while PET radiopharmaceuticals dominated brain protocols (74-89% ED).
- Scan length and patient positioning significantly impacted CT ED in 18F-FDG PET/CT (p < 0.01).
Conclusions:
- The mean effective dose for body and brain PET/CT protocols varies significantly based on radiopharmaceutical and scan type.
- CT is the main source of radiation dose for body PET/CT, whereas the PET radiopharmaceutical is the primary contributor for brain PET/CT.
- Optimizing scan parameters and radiopharmaceutical selection can help manage patient effective dose in PET/CT imaging.