Related Experiment Video
Updated: Aug 15, 2026

Ex Vivo Infection of Murine Epidermis with Herpes Simplex Virus Type 1
Published on: August 24, 2015
Effectiveness of a procedure for disinfecting the vaginal photoplethysmograph contaminated with herpes simplex virus
P J Morokoff1, L S Myers, J Hay
1Department of Psychology, University of Rhode Island, Kingston 02881.
Abstract:
To test the effectiveness of a disinfecting procedure involving 2% glutaraldehyde, a vaginal photoplethysmograph was contaminated with a known amount of herpes simplex type 2 (HSV-2). The vaginal photoplethysmograph was then put through the disinfection procedure. Two virus solutions were tested, one designed to approximate the concentration found in a naturally occurring infection (low inoculum), the other with 100 times more virus (high inoculum). Varying lengths of exposure to glutaraldehyde were tested. Results of assays for the virus after the device was disinfected showed that no measurable infectious virus remained even at the shortest exposure to glutaraldehyde, 1 min. A second experiment was conducted in which glutaraldehyde was added directly to a virus solution. Results confirmed those of the first experiment.