Harmonizing standardized uptake value recovery between two PET/CT systems from different manufacturers when using
Ian S Armstrong1, Katharine E Thomson, Lisa M Rowley
1aDepartment of Nuclear Medicine, Central Manchester University Hospitals, Manchester bDepartment of Medical Physics, Derriford Hospital, Plymouth Hospitals NHS Trust, Plymouth cDepartment of Nuclear Medicine, University Hospitals of Leicester NHS Trust, Leicester dDepartment of Radiation Physics and Protection, Oxford University Hospitals NHS Foundation Trust eDepartment of Oncology, University of Oxford, Oxford, UK.
Abstract:
PET iterative reconstruction algorithms with resolution modelling (RM) can be used to improve spatial resolution in the images. However, RM has a significant impact on quantification, which raises issues for harmonization across multicentre networks or collaborations. This investigation compared quantification from two modern time-of-flight (TOF) PET/CT systems from different manufacturers with RM with the intention to harmonize recovery. Images of a National Electrical Manufacturers Association image quality phantom with a sphere-to-background concentration ratio of 4 : 1 were acquired on a GE Discovery 710 and a Siemens Biograph mCT and reconstructed with RM and TOF. Voxel dimensions and image noise (background coefficient of variation) were matched. One to five iterations were used with 2 and 4 mm Gaussian filters. Mean and maximum contrast recovery (CR) were measured for the 10, 13, 17 and 22 mm hot phantom spheres. Notable differences in CR for images reconstructed with matched reconstruction parameters were observed between the scanners. A set of parameters was found that reduced differences in CR between scanners. Using these parameters, relative differences for the Biograph compared with the Discovery were -8.1, -3.7, +7 and +0.7% for mean CR and -23.1, -6.9, +9.1 and +0.9% for maximum CR in the 10, 13, 17 and 22 mm spheres, respectively. This study has used a technique of harmonizing standardized uptake value recovery on PET/CT systems from different vendors with advanced reconstructions including TOF and RM using phantom data. Considerable quantitative differences may occur in images, which highlights the need to apply methods such as those used in this work for multicentre studies.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...