[Cemented polyethylene insert in a well fixed Total knee replacement. Case report].
F E García-Dobarganes Barlow1,2,3, J Negrete-Corona4, A A Díaz-Tovar4
1División de Cirugía del Hospital Ángeles Querétaro. Calle Bernardino del Razo 21, Ensueño, 76178 Santiago de Querétaro, Qro. México.
Acta Ortopedica Mexicana
|April 28, 2017
Summary
A novel technique for total knee replacement revision in a 94-year-old patient involved cementing a new polyethylene insert onto the existing tibial baseplate. This approach successfully resolved pain and instability, restoring good function after two years.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Geriatric Medicine
Background:
- Osteoarthritis necessitates total knee replacement (TKR).
- Polyethylene wear is a common complication in TKR, potentially leading to revision surgery.
- Availability of specific implant components can be limited in revision cases.
Observation:
- A 94-year-old patient presented with pain and instability 14 years after right TKR due to osteoarthritis.
- Radiographs revealed polyethylene wear but no loosening of the metal tibial components.
- The original polyethylene insert was no longer manufactured.
Findings:
- A new polyethylene insert was cemented onto the native tibial baseplate, adapting a technique from total hip revisions.
- Two-year follow-up demonstrated the patient was asymptomatic with good functional recovery.
- No signs of radiolucency or loosening were observed at the tibial component interface.
Implications:
- Cementing a new polyethylene insert onto an existing tibial baseplate is a viable salvage option for TKR revision when original components are unavailable.
- This technique may offer a successful alternative to complete component revision, potentially reducing surgical complexity and cost.
- Further investigation into this salvage technique for polyethylene wear in TKR is warranted.