Related Experiment Video
Updated: Mar 3, 2026

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Asymmetric changes in circulating insulin levels after an increase compared with a reduction in insulin pump basal
S A McAuley1,2, G M Ward2,3, J C Horsburgh1
1Department of Medicine, University of Melbourne, Melbourne, Australia.
Aims:
To investigate circulating insulin profiles after a clinically relevant insulin pump basal rate increase vs a reduction, and the associated glucose responses.
Methods:
A cohort of 12 adults with Type 1 diabetes undertook this two-stage university hospital study using Accu-Chek pumps (Roche Diagnostics, Mannheim, Germany) and insulin aspart. An insulin basal rate change of 0.2 unit/h (increase in first stage, reduction in second stage) was implemented at ~09:30 h, after a single overnight basal rate (without bolus insulin), while fasting participants rested. Frequent venous samples for the assessment of plasma free insulin, glucose and cortisol were collected from 60 min before until 300 min after rate change. The primary outcome was time to steady-state insulin.
Results:
The 0.2-unit/h rate change represented a mean ± sd alteration of 23 ± 6%. After the rate increase, the median (interquartile range) times to 80% and 90% steady-state insulin were 170 (45) min and 197 (87) min, respectively. By contrast, after rate reduction, 80% steady-state insulin was not achieved. After the rate increase, mean ± se insulin levels increased by 4.3 ± 3.1%, 12.0 ± 2.9% and 25.6 ± 2.6% at 60, 120 and 300 min, respectively (with no significant difference until 180 min). After the rate reduction, insulin decreased by 8.3 ± 3.0% at 300 min (with no significant difference until 300 min). After rate reduction, glucose levels paradoxically declined by 17.4 ± 3.7% after 300 min; cortisol levels also fell during observation (P = 0.0003).
Conclusions:
The time to circulating insulin change after a 0.2-unit/h basal rate change was substantial, and was greater after a reduction than after an increase. Counter-regulatory hormone circadian variation may affect glycaemia when implementing minor changes at low basal rates. Both direction of basal rate change, and time of day, warrant consideration when anticipating the clinical effects of basal rate changes.
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are...
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Diabetes Mellitus: Type 2 and Gestational