Telescopic Corsair in 5F Guidion for Retrograde Recanalization of Complex Chronic Total Occlusions (CTOs)
Pieter-Jan Palmers1,2, Joren Maeremans1,3, Markus Meyer-Gessner4
1Department of Cardiology, Hospital East Limburg, Genk, Belgium.
Abstract:
BACKGROUND Retrograde advancement of microcatheters through septal/epicardial connectors can be challenging. Although several tricks might help to do so (e.g., balloon trap of retro wire in second guiding, balloon trap of retro wire in native coronary artery, and use of antegrade extension to approximate the antegrade conduit to the retrograde gear), these tricks cannot always be applied, especially in patients with poor access. Also, puncturing, knuckling, and crossing of the distal CTO cap (or the aorta as described in 1 of the cases) sometimes needs a lot of backup of the microcatheter. CASE REPORT We describe 3 cases in which we used a novel telescopic technique with 5F Guidion (IMDS®) supported retrograde Corsair (Asahi®) advancement in complex CTO lesions. CONCLUSIONS The telescopic Corsair in 5F Guidion may offer the support needed to end successfully in these situations.