Effect of Pressurized Metered Dose Inhaler Spray Characteristics and Particle Size Distribution on Drug Delivery
Morteza Yousefi1, Kiao Inthavong1, Jiyuan Tu1
1School of Engineering, RMIT University , Bundoora, Australia .
Background:
A key issue in pulmonary drug delivery is improvement of the delivery device for effective and targeted treatment. Pressurized metered dose inhalers (pMDIs) are the most popular aerosol therapy device for treating lung diseases. This article studies the effect of spray characteristics: injection velocity, spray cone angle, particle size distribution (PSD), and its mass median aerodynamic diameter (MMAD) on drug delivery.
Methods:
An idealized oral airway geometry, extending from mouth to the main bronchus, was connected to a pMDI device. Inhalation flow rates of 15, 30, and 60 L/min were used and drug particle tracking was a one-way coupled Lagrangian model.
Results:
The results showed that most particles deposited in the pharynx, where the airway has a reduced cross-sectional area. Particle deposition generally decreased with initial spray velocity and with increased spray cone angle for 30 and 60 L/min flow rates. However, for 15 L/min flow rate, the deposition increased slightly with an increase in the spray velocity and cone angle. The effect of spray cone angle was more significant than the initial spray velocity on particle deposition. When the MMAD of a PSD was reduced, the deposition efficiency also reduces, suggesting greater rates of particle entry into the lung. The deposition rate showed negligible change when the MMAD was more than 8 μm.
Conclusion:
Spray injection angle and velocity change the drug delivery efficacy; however, the efficiency shows more sensitivity to the injection angle. The 30 L/min airflow rate delivers spray particles to the lung more efficiently than 15 and 60 L/min airflow rate, and reducing MMAD can help increase drug delivery to the lung.
Related Concept Videos
Inhaled Medications
Factors Affecting Dissolution: Particle Size and Effective Surface Area
Modified-Release Drug Delivery Systems: Influencing Factors
Drug Delivery: Miscellaneous Routes
Oral inhalation and nasal sprays swiftly transfer drugs across the respiratory epithelium's mucosal layer. Inhaled glucocorticoids and bronchodilators directly target lung conditions such as asthma, while fluticasone nasal spray mitigates allergic rhinitis.
Transdermal patches transport drugs...
Modified-Release Drug Delivery Systems: Drug Release Characteristics
Pharmaceutical Alternatives: Polymorphic Form-Related and Particle Size-Related Therapeutic Nonequivalence