Impact of eye-tracking technology on OCT-angiography imaging quality in age-related macular degeneration
J L Lauermann1, M Treder1, P Heiduschka1
1Department of Ophthalmology, University of Muenster Medical Center, Domagkstrasse 15, 48149, Muenster, Germany.
Objective:
To evaluate the impact of eye-tracking (ET) technology on optical coherence tomography angiography (OCT-A) image quality and manifestation of motion artifacts in patients with age-related macular degeneration (AMD).
Methods:
In a prospective trial, multimodal retinal imaging including OCT-A was performed in 30 patients (78.97 ± 9.7 years) affected by different stages of AMD. Central 3 × 3 mm2 OCT-A imaging was performed four times consecutively in each patient, twice with active, and twice with inactive ET. Parameters for image evaluation were signal strength index (SSI), variability of foveal vessel density (VD), acquisition time, presence of motion artifacts caused by eye movement (blink lines, displacement) and by software correction of eye movement (quilting, stretch artifacts, vessel doubling). Images were evaluated by two independent readers with subsequent senior reader arbitration for presence of artifacts, and an OCT-A motion artifact score (MAS) was calculated.
Results:
Eight patients had early and eight patients had intermediate stages of AMD. Four patients had an atrophic late stage and ten patients an exudative stage of the disease. SSI was 53.55 with inactive and 57.18 with active ET (p = 0.0005). Coefficients of variability of VD between the first and second measurement were 8.9% with inactive and 5.7% with active ET. Mean image acquisition time was 15.97 s (active ET: 22.88 s, p < 0.001). Presence of motion artifacts was significantly higher with inactive ET (mean MAS 3.27 vs. 1.93; p < 0.0001). MAS correlated with AMD disease stage [p = 0.0031 (inactive ET) and p < 0.0001 (active ET)] and with SSI (p = 0.0072 and p = 0.0006).
Conclusions:
In patients with AMD, active ET technology offers an improved image quality in OCT-A imaging regarding presence of motion artifacts at the expense of higher acquisition time.