Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
Heat Generation on Implant Surface During Abutment Preparation at Different Elapsed Time Intervals
Abdullah Al-Keraidis1, Khalil Aleisa, Ziad Nawaf Al-Dwairi
1*Consultant Prosthodontist, Dental Center, King Saud Medical City, Riyadh, Saudi Arabia. †Professor, Department of Prosthetic Dental Sciences, College of Dentistry, King Saud University, Riyadh, Saudi Arabia. ‡Professor, Department of Prosthodontics, Faculty of Dentistry, Jordan University of Science and Technology, Irbid, Jordan. §Professor and Consultant Prosthodontist (deceased), Department of Prosthetic Dental Sciences, College of Dentistry, King Saud University, Riyadh, Saudi Arabia. ¶Professor and Chairman, Department of Dentistry, National Yang-Ming University, Taipei, Taiwan. ‖Course Director, Dentistry, Warwick Medical School, University of Warwick, Warwick, England. #Professor, Dental Materials Unit, Clinic for Fixed and Removable Prosthodontics and Dental Materials Science, Center for Dental and Oral Medicine, University of Zurich, Zurich, Switzerland.
Purpose:
The purpose of this study was to evaluate heat generation at the implant surface caused by abutment preparation using a diamond bur in a high-speed dental turbine in vitro at 2 different water-coolant temperatures.
Materials And Methods:
Thirty-two titanium-alloy abutments were connected to a titanium-alloy implant embedded in an acrylic resin placed within a water bath at a controlled temperature of 37°C. The specimens were equally distributed into 2 groups (16 each). Group 1: the temperature was maintained at 20 ± 1°C; and group 2: the temperature was maintained at 32 ± 1°C. Each abutment was prepared in the axial plane for 1 minute and in the occlusal plane for 1 minute. The temperature of the heat generated from abutment preparation was recorded and measured at 3 distinct time intervals.
Results:
Water-coolant temperature (20°C vs 32°C) had a statistically significant effect on the implant's temperature change during preparation of the abutment (P < 0.0001).
Conclusion:
The use of water-coolant temperature of 20 ± 1°C during preparation of the implant abutment decreased the temperature recorded at the implant surface to 34.46°C, whereas the coolant temperature of 32 ± 1°C increased the implant surface temperature to 40.94°C.