Some Lessons Learned From a Comparison Between Sedimentation Velocity Analytical Ultracentrifugation and Size
Aditya V Gandhi1, Mark R Pothecary2, David L Bain1
1Department of Pharmaceutical Sciences, University of Colorado Anschutz Medical Campus, Aurora, Colorado 80045.
Abstract:
There are numerous problems with size exclusion chromatography (SEC), which often lead to inaccuracies in protein aggregate characterization. Hence, this study tested sedimentation velocity analytical ultracentrifugation (SV-AUC) as an orthogonal tool to SEC for quantifying the monomer and aggregates in intravenous immunoglobulin (IVIg) formulations. IVIg samples were subjected to agitation stress and analyzed using SEC mobile phases composed of 200 mM sodium phosphate (pH 7.0) with 0, 50, 100, 200, or 400 mM of NaCl. Surprisingly, 400 mM of NaCl was required in the mobile phase to attain complete protein recovery from the SEC column. Significant discrepancies between SEC and SV-AUC were reported when SEC analysis was performed using suboptimal concentrations (e.g., 0, 50, 100, and 200 mM) of NaCl in the mobile phase. The continuous sedimentation coefficient distributions obtained with SV-AUC resolved the high molecular weight species, whereas with SEC the high molecular weight species eluted as a single peak. Only with the orthogonal use of SV-AUC, we were able to develop a robust SEC method for accurate quantitation of monomer and aggregates in unagitated and agitated IVIg samples. Additionally, this article describes a modification to an existing method of quantitating insoluble aggregates from SV-AUC boundary data.
Related Concept Videos
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Centrifugation
Two-dimensional Gel Electrophoresis
The first dimension separation uses the isoelectric focusing or IEF technique performed on immobilized pH gradient (IPG) strips that separate proteins according to their isoelectric points.
Biological samples, such...
SDS-PAGE
A variation of gel electrophoresis, termed polyacrylamide gel electrophoresis (PAGE), is commonly used for separating proteins according to their molecular size by passing them through a polyacrylamide gel. Because of the varying charges associated with amino acid side chains, PAGE can be used to separate intact...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Types Of Column Chromatography
Gel Filtration Chromatography
When the...