Related Experiment Video
Updated: Mar 2, 2026

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Alkaline phosphatase binds tenaciously to titanium; implications for biological surface evaluation following bone
J P Mansell1, A I Shiel1, C Harwood2
1Department of Biological, Biomedical & Analytical Sciences, University of the West of England, Frenchay Campus, Coldharbour Lane, Bristol BS16 1QY, UK.
Abstract:
Enhancing the performance and longevity of titanium (Ti) implants continues to be a significant developmental theme in contemporary biomaterials design. Our specific focus pertains to the surface functionalisation of Ti using the bioactive lipid, lysophosphatidic acid (LPA) and certain phosphatase-resistant analogues of LPA. Coating survivorship to a plethora of testing regimens is required to align with due regulatory process before novel biomaterials can enter clinical trials. One of the key acceptance criteria is coating retention to the physical stresses experienced during implantation. In assessing coating stability to insertion into porcine bone we found that a subsequent in vitro assessment to confirm coating persistence was masked by abundant alkaline phosphatase (ALP) contamination adsorbed to the metal surface. Herein we report that ALP can bind to Ti in a matter of minutes by simply immersing Ti samples in aqueous solutions of the enzyme. We strongly discourage the in vitro monitoring of osteoblast and stromal cell ALP expression when assessing bioactive coating survivorship following Ti implant retrieval form native bone tissue.