Related Experiment Video
Updated: Mar 2, 2026

Deep Vein Thrombosis Induced by Stasis in Mice Monitored by High Frequency Ultrasonography
Published on: April 13, 2018
Design of a HIFU array for the treatment of deep venous thrombosis: a simulation study
Petr Smirnov1,2, Kullervo Hynynen1,2
1Sunnybrook Research Institute, Physical Sciences Platform, Toronto, Canada.
Abstract:
Deep venous thrombosis of the iliofemoral veins is a common and morbid disease, with the recommended interventional treatment carrying a high risk of hemorrhaging and complications. High intensity focused ultrasound delivered with a single element transducer has been shown to successfully precipitate thrombolysis non-invasively in vitro and in vivo. However, in all previous studies damage to the veins or surrounding tissue has been observed. Using a simulation model of the human thigh, this study investigated whether a phased array device could overcome the large focal region limitations faced by single transducer treatment devices. Effects of the size, shape and frequency of the array on its focal region were considered. It was found that a [Formula: see text] spaced array of 7680 elements operating at 500 kHz could consistently focus to a region fully contained within the femoral vein. Furthermore, it is possible to reduce the number of elements required by building arrays operating at lower frequencies. The results suggest that phased transducer arrays hold potential for developing a safe, non-invasive treatment of thrombolysis.
More Related Videos
Related Concept Videos
Venous Thrombosis III: Interprofessional Care
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis I: Introduction