Validation of 3D-reconstructed computed tomography images using OsiriX® software for pre-transcatheter aortic valve
Marlena Gumsheimer1, Stefan Stortecky2, Brigitta Gahl1
1Department of Cardiac and Vascular Surgery, Inselspital University Hospital Bern, University of Bern, Bern, Switzerland.
Objectives:
We report validation of OsiriX® -an image processing freeware-to measure multi-slice computed tomography-derived annulus diameters for preprocedural transcatheter aortic valve implantation planning.
Methods:
A total of 137 patients (82 ± 6.5 years, 42.3% male, logistic EuroSCORE 24.1 ± 14.2%) with severe aortic stenosis at high surgical risk underwent transcatheter aortic valve implantation assessment: transoesophageal echocardiography, angiography and multi-slice computed tomography. Retrospectively, 3D multi-slice computed tomography reconstructions were generated using OsiriX and the reliability evaluated regarding inter- and intraobserver variability, intermodality correlation and estimation of the clinical impact on transcatheter aortic valve implantation sizing.
Results:
Reliability of the novel OsiriX software was high with an interobserver mean difference of 0.6 ± 1.4 mm and intraclass correlation of absolute agreement of 0.84 (95% confidence interval 0.74-0.90). The intermodality accuracy between OsiriX measurements and conventional 2D computed tomography reconstructions, transoesophageal echocardiography and angiography revealed significantly larger sizing with OsiriX, with a mean difference to 2D computed tomography of 0.4 ± 2.2 mm, which would have changed valve sizing in 38% of patients. In 28%, a larger size would have been chosen, and this correlated highly with the occurrence of postoperative severe aortic regurgitation (P < 0.001).
Conclusions:
While OsiriX measurements are an accurate and reproducible assessment of the aortic annulus, there are distinct and clinically relevant differences in aortic annulus dimensions between OsiriX measurements and previously standard imaging modalities. Sizing with OsiriX resulted in a larger perimeter compared with conventional 2D imaging. Careful assessment of valve size will take into account multiple imaging modalities.
More Related Videos
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
07:46Author Spotlight: Revolutionizing Remote Surgery with Augmented Reality and Robotics for Enhanced Precision and Accessibility
Published on: August 9, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT