Related Experiment Video
Updated: Mar 2, 2026

Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
Rapid-deployment aortic valve replacement versus standard bioprosthesis implantation
Enrico Ferrari1,2,3, Christelle Roduit2, Pauline Salamin2
1Cardiocentro Ticino Foundation, Cardiac Surgery Unit, Lugano, Switzerland.
The Intuity valve demonstrated lower transvalvular gradients compared to the Perimount Magna bioprosthesis in patients undergoing aortic valve replacement. Careful assessment of paravalvular leaks with the Intuity valve during surgery is recommended.
Area of Science:
- Cardiovascular Surgery
- Bioprosthetic Valves
- Hemodynamics
Background:
- Aortic valve replacement is a common procedure to treat severe aortic stenosis.
- Bioprosthetic valves are frequently used, with ongoing development of new designs for improved outcomes.
- Comparing the hemodynamic performance and safety of different valve types is crucial for clinical decision-making.
Purpose of the Study:
- To compare the 1-year clinical outcomes and hemodynamic results of the rapid-deployment Intuity valve and the Perimount Magna bioprosthesis.
- To evaluate the safety and efficacy of these two aortic bioprostheses in matched patient populations.
Main Methods:
- A retrospective comparison of 32 patients who received the Intuity valve versus a matched cohort receiving the Perimount valve.
- Clinical data collection and echocardiographic assessment at 1-year follow-up were performed.
- Key outcomes included surgical times, mortality, pacemaker implantation, hospital stay, and hemodynamic parameters (gradients, regurgitation).
Main Results:
- Shorter cross-clamp, cardiopulmonary bypass, and surgical times were observed with the Intuity valve.
- The Intuity valve group showed significantly lower mean gradients at discharge and 1-year follow-up compared to the Perimount group.
- While hospital mortality was zero for Intuity versus 3% for Perimount, some cases of aortic insufficiency were noted with the Intuity valve.
Conclusions:
- The Intuity valve exhibits favorable hemodynamic performance with lower gradients compared to the Perimount valve at similar mean valve sizes.
- Potential paravalvular leaks associated with Intuity valve implantation require vigilant intraoperative assessment and management.