Dose perturbations due to contrast medium and air in MammoSite® treatment: An experimental and Monte Carlo study
C-W Cheng1, R Mitra2, X Allen Li3
1Arizona Oncology Associates, 2625 N. Craycroft Road, Suite 100, Tucson, Arizona 85712.
Abstract:
In the management of early breast cancer, a partial breast irradiation technique called MammoSite® (Proxima Therapeutic Inc., Alpharetta, GA) has been advocated in recent years. In MammoSite, a balloon implanted at the surgical cavity during tumor excision is filled with a radio-opaque solution, and radiation is delivered via a high dose rate brachytherapy source situated at the center of the balloon. Frequently air may be introduced during placement of the balloon and/or injection of the contrast solution into the balloon. The purpose of this work is to quantify as well as to understand dose perturbations due to the presence of a high-Z contrast medium and/or an air bubble with measurements and Monte Carlo calculations. In addition, the measured dose distribution is compared with that obtained from a commercial treatment planning system (Nucletron PLATO system). For a balloon diameter of 42 mm, the dose variation as a function of distance from the balloon surface is measured for various concentrations of a radio-opaque solution (in the range 5%-25% by volume) with a small volume parallel plate ion chamber and a micro-diode detector placed perpendicular to the balloon axis. Monte Carlo simulations are performed to provide a basic understanding of the interaction mechanism and the magnitude of dose perturbation at the interface near balloon surface. Our results show that the radio-opaque concentration produces dose perturbation up to 6%. The dose perturbation occurs mostly within the distances <1mm from the balloon surface. The Plato system that does not include heterogeneity correction may be sufficient for dose planning at distances ⩾10mm from the balloon surface for the iodine concentrations used in the MammoSite procedures. The dose enhancement effect near the balloon surface (<1mm) due to the higher iodine concentration is not correctly predicted by the Plato system. The dose near the balloon surface may be increased by 0.5%percm3 of air. Monte Carlo simulation suggests that the interface effect (enhanced dose near surface) is primarily due to Compton electrons of short range (<0.5mm). For more accurate dosimetry in MammoSite delivery, the dose perturbation due to the presence of a radio-opaque contrast medium and air bubbles should be considered in a brachytherapy planning system.
More Related Videos
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
06:24Dynamic Contrast Enhanced Magnetic Resonance Imaging of an Orthotopic Pancreatic Cancer Mouse Model
Published on: April 18, 2015
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Magnetic Resonance Imaging