Numerical Evaluation of Image Contrast for Thicker and Thinner Objects among Current Intraoral Digital Imaging
Oyunbat Dashpuntsag1, Midori Yoshida1, Ryosuke Kasai2
1Department of Oral and Maxillofacial Radiology, Tokushima University, Tokushima, Japan.
Abstract:
The purpose is to evaluate the performance of current intraoral digital detectors in detail using a precise phantom and new method. Two aluminum step wedges in 0.5 mm steps were exposed by two photostimulable phosphor plate (PSP) systems-one with automatic exposure compensation (AEC) and the other without AEC-and a CCD sensor. Images were obtained with 3 doses at 60 kV. The effect of metallic material also was evaluated. The contrast-to-noise ratio (CNR) for thinner steps and the low contrast value (LCV) for thicker steps were obtained. The CCD system was the best under all conditions (P < 0.001), although the Gray value was sensitive to the dose, and the Gray value-dose relation varied greatly. The PSP system with AEC was superior to that without AEC for the LCV (P < 0.001) but was inferior to it regarding the CNR (P < 0.001). CNR and LCV in the PSP system without AEC were not affected by the metallic plate. Intraoral digital imaging systems should be chosen according to their diagnostic purpose. PSP system with AEC may be the best for detecting molar proximal caries, whereas the PSP system without AEC may be better for evaluating small bone regeneration in periodontal disease. The CCD system provided the best performance.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies III: Computed Tomography
X-ray Imaging
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...