Customized Hinged Covered Metallic Stents for the Treatment of Benign Main Bronchial Stenosis
Xinwei Han1, Quazi Al-Tariq2, Yanle Zhao2
1Department of Interventional Radiology, First Affiliated Hospital of Zhengzhou University, Zhengzhou, Henan, China.
Background:
To address the limitations of silicone stents, we designed a hinged self-expandable covered metallic stent. The aim of this study was to evaluate the safety and efficacy of the customized stents in clinical applications.
Methods:
This was a retrospective analysis. Under conscious sedation and local anesthesia, the stents were implanted or removed by interventional radiologists, with fluoroscopic guidance.
Results:
Of 24 patients with benign main bronchial stenosis, stents were successfully placed in 21 (87.5%). The low-pressure balloon before dilation failed in 1 case (4.17%) of left main bronchial cicatricial stenosis. In 2 other cases (8.33%), stent placement was abandoned. Stents were successfully removed between 29 and 103 days after the procedure. After stent removal, the follow-up lasted for at least 12 months. Restenosis occurred only in 1 case (4.55%) owing to bronchial collapse 3 days after stent removal. Dyspnea occurred in another case (4.55%) at 2 months after retrieval; recurrence was confirmed using bronchoscopy, leading to a left pneumonectomy.
Conclusions:
The described procedure is safe and easy to be performed and avoids the use of intubation, bronchoscopy, and general anesthesia.
Related Concept Videos
Mitral Stenosis III: Medical Management
COPD: Management Using Bronchodilators and Corticosteroids
Mitral Stenosis II: Clinical features and Diagnostic Tests
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Mitral Stenosis IV: Nursing Management