Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Suitability of open-field autorefractors as pupillometers and instrument design effects
Carles Otero1, Mikel Aldaba1, Oriol Ferrer1
1Davalor Research Center, Technical University of Catalonia, Terrassa 08222, Spain.
Aim:
To determine the agreement and repeatability of the pupil measurement obtained with VIP-200 (Neuroptics), PowerRef II (Plusoptix), WAM-5500 (Grand Seiko) and study the effects of instrument design on pupillometry.
Methods:
Forty patients were measured twice in low, mid and high mesopic. Repeatability was analyzed with the within-subject standard deviation (Sw) and paired t-tests. Agreement was studied with Bland-Altman plots and repeated measures ANOVA. Instrument design analysis consisted on measuring pupil size with PowerRef II simulating monocular and binocular conditions as well as with proximity cues and without proximity cues.
Results:
The mean difference (±standard deviation) between test-retest for low, mid and high mesopic conditions were, respectively: -0.09 (±0.16), -0.05 (±0.18) and -0.08 (±0.23) mm for Neuroptics, -0.05 (±0.17), -0.12 (±0.23) and -0.17 (±0.34) mm for WAM-5500, -0.04 (±0.27), -0.13 (±0.37) and -0.11 (±0.28) mm for PowerRef II. Regarding agreement with Neuroptics, the mean difference for low, mid and high mesopic conditions were, respectively: -0.48 (±0.35), -0.83 (±0.52) and -0.38 (±0.56) mm for WAM-5500, -0.28 (±0.56), -0.70 (±0.55) and -0.61 (±0.54) mm for PowerRef II. The mean difference of binocular minus monocular pupil measurements was: -0.83 (±0.87) mm; and with proximity cues minus without proximity cues was: -0.30 (±0.77) mm.
Conclusion:
All the instruments show similar repeat-ability. In all illumination conditions, agreement of Neuroptics with WAM-5500 and PowerRef II is not good enough, which can be partially induced due to their open field design.