An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Effect of Implant Diameter And Alloy on Peri-Implant Strain: An In Vitro Quantitative Strain Analysis
Dinusha Goonawardhana1, Roy Judge1, Joseph Palamara1
1Melbourne University.
Abstract:
This study aimed to evaluate the impact of implant diameter and alloy on peri-implant strains. Three implant types were investigated: regular commercially pure titanium (RcpTi), narrow commercially pure titanium (NcpTi) and narrow titanium-zirconium alloy (NTi-Zr) implants. Strain gauge models were fabricated from epoxy resin. All the implants were loaded in identical fashion. The greatest peri-implant strains were recorded around NTi-Zr implants, followed by NcpTi implants. The RcpTi implants were associated with the least strain values. The difference between the implant types was significant. Therefore, it can be concluded that the implant diameter and material can influence the peri-implant strain magnitude.
More Related Videos
06:36Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
03:10Pilot In Vitro Study to Assess Cleaning Ability and Effects of Different Decontamination Methods on Implant Surfaces
Published on: November 21, 2025
Related Concept Videos
Measurements of Strain
Three-Dimensional Analysis of Strain
Stress-Strain Diagram
True Stress and True Strain
In contrast, true stress offers a more precise portrayal. It is computed by dividing the...
Transformation of Plane Strain
Under plane strain conditions, typical for members where one dimension significantly exceeds the others, deformations and resultant strains are...