Related Experiment Video
Updated: Mar 2, 2026

Author Spotlight: Advancing Awake Nasotracheal Intubation with Flexible Video Rhino-Laryngoscopes
Published on: August 2, 2024
Comparison of endotracheal intubation time in neutral position between C-Mac® and Airtraq® laryngoscopes: A
Syed Moied Ahmed1, Kashmiri Doley1, Manazir Athar1
1Department of Anaesthesiology and Critical Care, Jawaharlal Nehru Medical College and Hospital, Aligarh Muslim University, Aligarh, Uttar Pradesh, India.
Background And Aims:
In the recent past, many novel devices such as AirTraq® and C-MAC® video laryngoscope (VL) have been introduced in an attempt to reduce anaesthetic morbidity and mortality associated with difficult intubation. In this study, we aimed to evaluate and compare C-MAC® VL with a standard Macintosh blade and the AirTraq® optical laryngoscope as a intubating devices with the patient's head in neutral position.
Methods:
Sixty American Society of Anesthesiologist Physical Status I-II patients were randomly assigned to be intubated with C-MAC® VL (Group CM; n = 30) or AirTraq® (Group AT; n = 30) in the neutral position, with or without the application of optimization manoeuvres. The primary outcomes of this study were the success rate and the time taken to intubate. Glottic view, ease of tracheal intubation and haemodynamic responses were considered as secondary end points.
Results:
The incidence of successful intubation was similar in both the groups (P = 1.00). However, the time for intubation was significantly less with C-MAC® VL (Group CM = 14.9 ± 12.89 s, Group AT = 26.3 ± 13.34 s; P = 0.0014). There was no significant difference between the two groups in terms of ease of intubation and glottic view. However, the haemodynamic perturbations were much less with C-MAC® VL.
Conclusion:
We conclude that both the devices were similar in visualising larynx in the neutral position with similar success rates of intubation. However, the C-MAC® VL was better with respect to intubation time and haemodynamic stability.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...