Validation of Repeated Endothelial Function Measurements Using EndoPAT in Stroke.
Aina S Hansen1, Jawad H Butt1, Sonja Holm-Yildiz1
1Department of Neurology, Stroke Unit, Neurovascular Research Unit, Herlev-Gentofte University Hospital, Herlev, Denmark.
Frontiers in Neurology
|May 19, 2017
Summary
Reliability of endothelial function (EF) measurements using pulse amplitude tonometry was assessed in stroke patients and healthy individuals. Day-to-day measurements showed better reliability than same-day tests, especially for Framingham RHI in stroke patients.
Area of Science:
- Vascular Medicine
- Cardiology
- Neurology
Background:
- Endothelial function (EF) is a potential prognostic marker for stroke.
- Personalized medicine approaches for stroke prevention may involve measuring pharmacological effects on EF.
- Assessing the reliability of EF measurements is crucial for clinical application.
Purpose of the Study:
- To evaluate the reliability of repeated endothelial function (EF) measurements using pulse amplitude tonometry.
- To compare measurement reliability in acute stroke patients versus healthy subjects.
- To determine the most reliable outcome measure for EF assessment.
Main Methods:
- Endothelial function (EF) was measured using the EndoPAT2000 device in 20 healthy volunteers and 21 stroke patients.
- Measurements included reactive hyperemia index (RHI), logarithm of RHI (lnRHI), and Framingham RHI (fRHI).
- Reliability was assessed through same-day (1.5 hours apart) and day-to-day (24 hours apart) repeated measurements.
Main Results:
- Healthy subjects showed fair to moderate same-day (ICC=0.29) and moderate day-to-day (ICC=0.52) reliability.
- Stroke patients demonstrated moderate same-day (ICC=0.40) and substantial day-to-day (ICC=0.62) reliability.
- Framingham RHI (fRHI) exhibited superior reliability compared to RHI and lnRHI.
Conclusions:
- Repeated fRHI measurements show moderate same-day and substantial day-to-day reliability in stroke patients.
- Day-to-day EF measurements are generally more reliable than same-day measurements across both groups.
- Potential carryover effects necessitate caution when considering same-day repeated testing for drug efficacy studies.