Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Instrument Calibration01:12

Instrument Calibration

916
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
916
Glassware Calibration01:11

Glassware Calibration

1.6K
Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
1.6K
Gas Chromatography: Types of Detectors-II01:19

Gas Chromatography: Types of Detectors-II

1.3K
In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
1.3K
Gas Chromatography: Types of Detectors-I01:21

Gas Chromatography: Types of Detectors-I

1.8K
There are different types of detectors used in gas chromatography, each with its own specific properties that make it suitable for detecting certain types of analytes. The most commonly used detectors in GC are thermal conductivity detector (TCD), flame ionization detector (FID), and electron capture detector (ECD).
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
1.8K
High-Performance Liquid Chromatography: Types of Detectors01:15

High-Performance Liquid Chromatography: Types of Detectors

1.9K
The role of the detectors in High-Performance Liquid Chromatography (HPLC) is to analyze the solutes as they exit from the chromatographic column. The detector recognizes the solute's property and generates corresponding electrical signals, which are converted into a readable graph of the detector's response versus elution time called a chromatogram at the computer. There are several types of HPLC detectors, each with its own advantages and limitations, depending on the analyte...
1.9K
Gas Chromatography: Overview of Detectors01:13

Gas Chromatography: Overview of Detectors

2.2K
Detectors in gas chromatography (GC) help identify and quantify the components of a mixture by translating chemical properties into measurable signals, which are displayed on a chromatogram. Detectors can be categorized into two main types: destructive and non-destructive.
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
2.2K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Sci-Fri PM: Delivery - 06: A generalized solution to the wide field array calibration method.

Medical physics·2017
Same author

Poster - Thurs Eve-32: Dose errors related to the treatment couch.

Medical physics·2017
Same author

Comparison of the Diagnostic Accuracy of DSC- and Dynamic Contrast-Enhanced MRI in the Preoperative Grading of Astrocytomas.

AJNR. American journal of neuroradiology·2015
Same author

Preoperative prognostic value of dynamic contrast-enhanced MRI-derived contrast transfer coefficient and plasma volume in patients with cerebral gliomas.

AJNR. American journal of neuroradiology·2014
Same author

Diagnostic accuracy of dynamic contrast-enhanced MR imaging using a phase-derived vascular input function in the preoperative grading of gliomas.

AJNR. American journal of neuroradiology·2012
Same author

The emerging role of IG-IMRT for palliative radiotherapy: a single-institution experience.

Current oncology (Toronto, Ont.)·2009

Related Experiment Video

Updated: Mar 2, 2026

Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers
10:29

Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers

Published on: March 21, 2016

12.9K

Poster - Thur Eve - 32: Water tank referenced calibration method for detector array devices.

C Foottit1, L Gerig1

  • 1The Ottawa Hospital Cancer Centre, Ottawa, ON.

Medical Physics
|May 19, 2017
PubMed
Summary

A new calibration method using water tank data improves detector array accuracy for radiation therapy. This approach enhances beam profile evaluation by reducing errors associated with standard calibration techniques.

Keywords:
CalibrationDetector arraysDosimetryLinear accelerators

More Related Videos

Thermal Measurement Techniques in Analytical Microfluidic Devices
08:29

Thermal Measurement Techniques in Analytical Microfluidic Devices

Published on: June 3, 2015

10.2K
Terahertz Microfluidic Sensing Using a Parallel-plate Waveguide Sensor
07:28

Terahertz Microfluidic Sensing Using a Parallel-plate Waveguide Sensor

Published on: August 30, 2012

11.2K

Related Experiment Videos

Last Updated: Mar 2, 2026

Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers
10:29

Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers

Published on: March 21, 2016

12.9K
Thermal Measurement Techniques in Analytical Microfluidic Devices
08:29

Thermal Measurement Techniques in Analytical Microfluidic Devices

Published on: June 3, 2015

10.2K
Terahertz Microfluidic Sensing Using a Parallel-plate Waveguide Sensor
07:28

Terahertz Microfluidic Sensing Using a Parallel-plate Waveguide Sensor

Published on: August 30, 2012

11.2K

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Radiotherapy Physics

Background:

  • Detector array devices like the I'mRT Matrixx are crucial for evaluating radiation therapy beam profiles.
  • Standard relative calibration of these arrays is sensitive to beam output variations.
  • An alternative calibration method referencing detector response to water tank data is proposed.

Purpose of the Study:

  • To introduce and validate a novel water tank reference calibration procedure for detector arrays.
  • To assess the accuracy and energy independence of the proposed calibration method.
  • To mitigate errors in detector array calibration for improved beam profile analysis.

Main Methods:

  • Measured Matrixx detector response at four cardinal angles for three devices.
  • Determined calibration factors by dividing water tank profiles by Matrixx response.
  • Calculated the geometric mean of orientations to estimate the calibration coefficient.

Main Results:

  • Reduced average deviation from water tank profiles from 0.4% to 0.1% after calibration.
  • Confirmed the energy independence of the proposed relative calibration method.
  • Mitigated differences between short Matrixx acquisitions and long water tank acquisitions through averaging.

Conclusions:

  • The proposed water tank reference calibration is effective for detector arrays.
  • This method mitigates compound errors by avoiding recursive algorithms.
  • The procedure offers direct relevance to commissioning beam data in radiotherapy.