X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Poster - Thur Eve - 43: Is faster always better? An evaluation of frame rate effects on continuous acquisition mode
S Peca1, D Granville2, D Brown1
1University of Calgary, Tom Baker Cancer Centre, Calgary, AB.
Introduction:
As radiotherapy moves towards intensity modulated arc therapy (arc-IMRT), there is a need for electronic portal imaging (EPID) to move towards continuous acquisition (cine) mode for dosimetric verification purposes. However, as the EPID resolution and frame rate (fps) increase, so does the computational burden of image processing. We investigated the reliability of cine mode EPID imaging in IMRT as a function of frame rate.
Methods:
We acquired EPID images continuously while running an IMRT plan (6MV photons, 150MU, dose rate = 300MU/min) with frame rates ranging from 1-12 fps, as well as a single integrated mode image. Each cine dataset was then averaged to form a single image, which was compared with the integrated mode image by means of a pixel-by-pixel absolute value subtraction.
Results:
Although a greater frame rate gave better agreement with the integrated mode image in all cases, the relative benefit diminished with increasing frame rate. In particular, for the IMRT plan delivered, there was little benefit of imaging faster than 6 fps, and virtually no benefit in increasing from 9 to 12 fps. In contrast, 12 fps produces twice the number of images as 6 fps which significantly increases the image processing and data storage burdens.
Conclusion:
Increasing frame rate in cine mode EPID imaging may be beneficial in some cases, but there is likely a threshold level above which no relevant additional information is obtained. Further research to determine the ideal frame rate for any particular IMRT or arc-IMRT plan is warranted.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Dose Size and Dosing Frequency: Determination Methods