SU-E-T-573: Quality and Deliverability of Intensity Modulated Neutron Radiotherapy (IMNRT) Plans.
R Spink1,2, J Burmeister1,2, T Bossenberger1,2
1Karmanos Cancer Center, Detroit, MI.
Medical Physics
|May 19, 2017
Summary
Intensity Modulated Neutron Radiotherapy (IMNRT) plan quality is limited by multi-leaf collimator speed. Limiting IMNRT plans to 25 segments ensures reasonable treatment times while meeting most TG-119 dose constraints.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Technology
Background:
- Intensity Modulated Neutron Radiotherapy (IMNRT) is a new clinical technique.
- The number of segments in IMNRT plans is restricted by the speed of the multi-leaf collimator (MLC).
- Guidelines for IMNRT plan complexity are needed to ensure quality and deliverability.
Purpose of the Study:
- To evaluate the quality and deliverability of static IMNRT treatment plans.
- To establish guidelines for the maximum number of segments per IMNRT plan.
- To assess compliance with TG-119 dose constraints for varying plan complexities.
Main Methods:
- Treatment plans were optimized using the Varian Eclipse Treatment Planning System (TPS).
- A 4MV photon beam served as a surrogate for neutron beam optimization.
- MLC segments were generated for three complexity levels: very limited (15-23), limited (24-31), and unlimited.
- Dose-volume histograms (DVHs) were analyzed for TG-119 compliance.
- Estimated delivery times were calculated based on known parameters.
Main Results:
- The prostate case met all TG-119 constraints across all complexity levels.
- Other cases failed to meet at least one constraint at certain complexity levels.
- The number of met constraints increased with complexity: 16 (very limited), 17 (limited), 19 (unlimited) out of 23 total.
- Mean delivery times were 34, 40, and 68 minutes for very limited, limited, and unlimited complexity, respectively.
Conclusions:
- Current MLC capabilities limit the quality of IMNRT plans.
- IMNRT plans should be restricted to approximately 25 segments.
- This segment limitation ensures a treatment time under 45 minutes.
- Most TG-119 dose constraints can be met even with a limited number of segments.