SU-E-T-118: Characterization of EBT3 Films in Photon and Proton Beams
J Sorriaux1,2,3,4, S Rossomme1,2,3,4, A Kacperek1,2,3,4
1Molecular Imaging Radiotherapy & Oncology, Catholic University of Louvain, Belgium.
Purpose:
To measure the calibration curves of EBT3 dosimetry films in photon and proton beams and to quantify the related uncertainties from one beam type to another.
Methods:
EBT3 Gafchromic films have similar properties than EBT2 with a symmetric construction and a matte polyester substrate to prevent Newton's ring artefacts. Films from a same batch were exposed in three different beam qualities, an Elekta SL25 6 MV photon beam, a 100 MeV 5×5cm2 proton beam delivered by pencil-beam scanning dedicated system from IBA and a 60 MeV fixed proton beam (2.5cm in diameter) at Clatterbridge Center for Oncology (CCO), UK. The films were read using an EPSON 10000 XL/PRO scanner. Film calibration curves were acquired for all modalities within a range of 0.05 to 20 Gy. Influence of increasing linear-energy transfer (LET) on film response was investigated by comparing dose measured by EBT3 to a silicon diode detector in depth for a fully-modulated beam using the CCO beam line (homogeneous dose with distal end at 3.1cm in water). A comprehensive uncertainty budget (reproducibility, uniformity'¦) was estimated on films irradiated by Elekta SL25.
Results:
The main source of uncertainty was the non-uniformity of the scanner response. By placing all the irradiated films at the center of the scanner, the uncertainty could be reduced from 5.8% to 1.9% (1 sigma). For all beams and energies, the calibration curves were matched within uncertainties. Along the fully-modulated depth dose curve, diode and EBT3 measurement were in a 4% agreement point-to-point, indicating films weak dependence with LET.
Conclusions:
The weak influence of LET, beam type and energy on film response as well as its small uncertainty make EBT3 suitable for relative dosimetry and a promising candidate for measuring correction factors (quality, recombination,'¦) for reference dosimetry with ion chambers of non-standard beams (e.g pencil-beam scanning proton-therapy). “This work is supported by the Walloon Region under the project name InVivoIGT, convention number 1017266.â€.
Related Concept Videos
Thomson's e/m Experiment
A particle with charge q, speed v, and mass m enters an area from the top, where the magnetic and electric fields are perpendicular both to the particle's motion and to one another. The magnetic...
Electron Behavior
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the nucleus have less energy,...
Electron Behavior
Electrons are negatively charged subatomic particles that are attracted to an orbit around the positively-charged nucleus of an atom. They reside in locations that are associated with energy levels called shells and are further organized into sub-shells and orbitals within each shell.
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...