SU-E-T-96: Energy Dependence of the New GafChromic- EBT3 Film's Dose Response-Curve
S Chiu-Tsao1,2,3, G Massillon-Jl1,2,3, I Domingo-Muñoz1,2,3
1Quality MediPhys LLC, Denville, NJ.
Purpose:
To study and compare the dose response curves of the new GafChromic EBT3 film for megavoltage and kilovoltage x-ray beams, with different spatial resolution.
Methods:
Two sets of EBT3 films (lot#A101711-02) were exposed to each x-ray beam (6MV, 15MV and 50kV) at 8 dose values (50-3200cGy). The megavoltage beams were calibrated per AAPM TG-51 protocol while the kilovoltage beam was calibrated following the TG-61 using an ionization chamber calibrated at NIST. Each film piece was scanned three consecutive times in the center of Epson 10000XL flatbed scanner in transmission mode, landscape orientation, 48-bit color at two separate spatial resolutions of 75 and 300 dpi. The data were analyzed using ImageJ and, for each scanned image, a region of interest (ROI) of 2×2cm2 at the field center was selected to obtain the mean pixel value with its standard deviation in the ROI. For each energy, dose value and spatial resolution, the average netOD and its associated uncertainty were determined. The Student's t-test was performed to evaluate the statistical differences between the netOD/dose values of the three energy modalities, with different color channels and spatial resolutions.
Results:
The dose response curves for the three energy modalities were compared in three color channels with 75 and 300dpi. Weak energy dependence was found. For doses above 100cGy, no statistical differences were observed between 6 and 15MV beams, regardless of spatial resolution. However, statistical differences were observed between 50kV and the megavoltage beams. The degree of energy dependence (from MV to 50kV) was found to be function of color channel, dose level and spatial resolution.
Conclusions:
The dose response curves for GafChromic EBT3 films were found to be weakly dependent on the energy of the photon beams from 6MV to 50kV. The degree of energy dependence varies with color channel, dose and spatial resolution. GafChromic EBT3 films were supplied by Ashland Corp. This work was partially supported by DGAPA-UNAM grant IN102610 and Conacyt Mexico grant 127409.
More Related Videos
11:51Engineering 'Golden' Fluorescence by Selective Pressure Incorporation of Non-canonical Amino Acids and Protein Analysis by Mass Spectrometry and Fluorescence
Published on: April 27, 2018
09:16Simultaneous ex vivo Functional Testing of Two Retinas by in vivo Electroretinogram System
Published on: May 6, 2015
Related Concept Videos
Dose-Response Relationship: Overview
Dose Response Curve: Conventional Versus Nonmonotonic
Pharmacokinetic–Pharmacodynamic Relationship: Dose to Pharmacological Effect
Pharmacokinetic–Pharmacodynamic Relationship: Intensity of Dose-Effect Relationship
Biological Effects of Radiation