Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
SU-E-T-105: The LET Dependence of Liquid Ionization Chambers (LICs) in High-LET Beams
S Tegami1,2,3,4, F Gomez1,2,3,4, D Gónzales-Castaño1,2,3,4
1Max Planck Institute for Nuclear Physics, Heidelberg, Germany.
Purpose:
LICs are novel detectors for radiotherapy: the higher density of the medium allows to build them with a smaller sensitive volume, making them appealing in particle therapy. With RBE varying along the depth dose curve (DDC) and with the rising interest in dose/LET-painting, verifying the LET is becoming more important. Nevertheless, while the LET distributions for different ionizing particles have been calculated, they have never been directly measured in realistic therapeutic beams. Our interest in LICs is based on the characterization of the beam quality in terms of LET. It has been shown in earlier works that the extrapolation of the linear portion of the voltage curve yields an intercept with the x-axis that depends on LET. The quantitative establishment of this method, however, depends on how accurately recombination effects are taken into account.
Methods:
Due to the higher density of charge carriers produced in the liquid, LICs have high recombination effects: general recombination effects, involving pairs belonging to different tracks (dose rate dependent), and initial recombination between ion-electron pairs belonging to the same incident particle event (LET dependent). To perform this study we propose a two-dimensional array of LICs, composed by a 16×8 matrix of 2×2 mm2 pixels, which gives a fine spatial resolution on the plane.
Results:
Voltage curves have been measured for proton, carbon and oxygen beams available at the HIT facility in Heidelberg for different energies and dose rates. After correcting the curves for general recombination losses using the Three Voltage Method, we have indications of dose rate independence, indicating successful correction.
Conclusions:
Further investigations are foreseen to quantify the LET dependence along the DDC, where different LET values are expected. A comparison with simulated dose averaged LET values will give quantitative information about 2D LET distributions for different beam species.
More Related Videos
08:31Sample Preparation and Experimental Design for In Situ Multi-Beam Transmission Electron Microscopy Irradiation Experiments
Published on: June 27, 2022
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
Related Concept Videos
Chemical Ionization (CI) Mass Spectrometry
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
High-Performance Liquid Chromatography: Types of Detectors
Atomic Emission Spectroscopy: Lab