SU-E-T-384: An Easy Use Anthropomorphic Phantom for Lung IMRT QA.
E Fan1, D Ellerbusch1, L Qin1
1University of Minnesota, Minneapolis, MN.
This study developed a heterogeneous phantom for validating lung intensity-modulated radiation therapy (IMRT) plans. The phantom, using anthropomorphic phantom slabs, proved effective for accurate, patient-specific quality assurance in clinical settings.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Quality Assurance
Background:
- Lung IMRT plans are calculated with CT heterogeneity correction.
- Patient-specific validation often uses homogeneous solid media, which may not accurately reflect patient anatomy.
- There is a need for realistic, heterogeneous phantoms for precise lung IMRT validation.
Purpose of the Study:
- To develop and evaluate a heterogeneous phantom for measurement-based validation of lung IMRT plans.
- To assess the accuracy of IMRT plan calculations against phantom measurements.
- To provide a practical tool for routine clinical quality assurance of lung IMRT.
Main Methods:
- Constructed two heterogeneous phantoms (Hetero-1, Hetero-2) using anthropomorphic RANDO phantom slabs and virtual water.
- Employed a 2D diode array detector (MapCHECK) and ion chamber for dose measurements.
- Generated two lung IMRT plans and compared point and planar doses with phantom measurements and Treatment Planning System (TPS) calculations.
Main Results:
- MapCHECK measurements for Hetero-1 showed high passing rates (93.5%-98.5%) under 3%/3mm Gamma criteria.
- Hetero-2 measurements yielded passing rates of 96.3%-98.5%, even with shifts including more lung material.
- Ion chamber measurements showed minimal differences (1.2%-1.71%) compared to TPS calculations, with TLD results within 98%-99% agreement.
Conclusions:
- An easily assembled heterogeneous phantom using anthropomorphic phantom slabs is feasible for lung IMRT QA.
- This phantom serves as a convenient and efficient tool for routine clinical validation of lung IMRT.
- The developed phantom enables accurate, patient-specific quality assurance for lung IMRT plans.