SU-E-T-386: Gamma Analysis of Normalized and Un-Normalized Dose Distributions
S Stojadinovic1, O Luo1, Q Bao1
1UT Southwestern Medical Center, Dallas, TX.
The gamma index method for radiation dose QA needs no normalization point for unique results. However, standard 3%/3 mm criteria are inadequate without normalization, requiring re-evaluation for accurate dose distribution agreement.
Area of Science:
- Medical Physics
- Radiation Oncology
- Quality Assurance
Background:
- Current gamma index methods in commercial QA software require selecting a normalization point.
- This normalization point choice leads to multiple possible solutions for dose distribution agreement.
- A unique solution is achieved by omitting the normalization point.
Purpose of the Study:
- To evaluate the impact of using a normalization point versus no normalization point in the gamma index method.
- To assess the agreement between calculated and delivered dose distributions using AAPM TG119 test cases and clinical data.
- To determine the suitability of common gamma index criteria (3%/3 mm) without normalization.
Main Methods:
- AAPM TG119 test cases were planned and delivered using Pinnacle 8.0m and a Varian 21EX linac (6 and 18 MV photons).
- Point and planar dose measurements were performed using ion chamber, film, and MatriXX.
- Gamma index analysis (3%/3 mm) was conducted with and without a normalization point on both benchmark and clinical data.
Main Results:
- With normalization, gamma passing rates for TG119 complex tests were high (95.8±1.6% for 6 MV, 95.6±1.0% for 18 MV).
- Without normalization, gamma passing rates significantly dropped (20.7±6.7% for 6 MV, 13.9±4.0% for 18 MV).
- Clinical data showed a similar trend, with passing rates decreasing from 98.9±0.7% to 33.4±13.1% without normalization.
Conclusions:
- The gamma index method yields a unique gamma passing rate solely when dose distributions are not normalized.
- The standard 3%/3 mm gamma criteria is an insufficient metric when no normalization point is used.
- Further investigation into appropriate gamma criteria for non-normalized dose comparisons is warranted.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Biological Effects of Radiation
Dose Response Curve: Conventional Versus Nonmonotonic
Normal Distribution
Introduction to Normal Distributions
Model-Independent Approaches for Pharmacokinetic Data: Noncompartmental Analysis
One important characteristic of noncompartmental analyses is that drug exposure increases proportionally with increasing doses. This...
z Scores and Area Under the Curve