SU-E-T-318: Source Position Localization for ICBT Tandem & Ovoid Applicators Utilizing Oncentra Brachy Applicator
Medical Physics
|May 19, 2017
Summary
Nucletron
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- Image artifacts in HDR cervical brachytherapy can hinder accurate dwell position localization.
- The Nucletron Oncentra Brachy Applicator Modeling plugin (AMp) offers a potential solution by using 3D CAD models for defining dwell positions.
Purpose of the Study:
- To compare the accuracy of source localization and dosimetric equivalence between clinical practice and the AMp for HDR cervical brachytherapy.
- To evaluate the impact of AMp-based localization on organs at risk (OARs) in a phantom study.
Main Methods:
- KVCT imaging of a water phantom with bladder and rectum surrogates and a Fletcher-type tandem and ovoid applicator.
- Catheter reconstruction using both clinical protocols (dwell position markers) and the AMp within Oncentra's Brachy TPS.
- Treatment planning to 600cGy at Point A, followed by spatial comparison of dwell positions and dosimetric analysis (D0.1cc, D1cc, D2cc) for surrogates.
Main Results:
- 80% of dwell positions localized by AMp agreed within 1mm of clinical protocol for a 25mm ovoid/30° tandem applicator.
- Maximum deviation between AMp and clinical protocol localization was 1.17mm.
- Dosimetric differences for OAR surrogates were minimal, with slight increases in D0.1cc, D1cc, and D2cc when using AMp.
Conclusions:
- Preliminary findings suggest that the AMp provides equivalent dosimetry compared to clinical protocols for HDR cervical brachytherapy in a simple phantom setup.
- Further investigation is ongoing with different ovoid sizes and a vaginal cylinder to validate these results.