SU-E-T-320: A New Verification Phantom for GYN Brachytherapy Applicators Using GafChromic - Films
S Gholami1,2,3,4, A S Meigooni1,2,3,4, Sr Mehdi Mahdavi1,2,3,4
1Cancer institution, radiotherapy group, Tehran.
A new phantom was developed for accurate brachytherapy dose verification in gynecological cancer treatment. This tool confirms radiation delivery accuracy for Cobalt-60 HDR systems, improving patient care.
Area of Science:
- Medical Physics
- Radiation Oncology
- Brachytherapy
Background:
- Brachytherapy is crucial for treating gynecological cancers like vaginal, cervical, and endometrial.
- Accurate dosimetry is essential for effective radiation therapy and minimizing side effects.
Purpose of the Study:
- To design, fabricate, and verify a novel phantom for precise dosimetric verification.
- To assess dose accuracy at small distances from gynecological (GYN) applicators using a Cobalt-60 High Dose Rate (HDR) system (GZP6).
Main Methods:
- A phantom was constructed from 90 Perspex slabs to house GYN applicators.
- GafChromic films were placed between slabs for dosimetry measurements with the GZP6 system.
- Film dosimetry results were compared against treatment planning software calculations.
Main Results:
- The study identified dose discrepancies of up to ±23% between planned and measured dosimetry.
- The developed phantom successfully confirmed the accuracy of radiation delivery for GYN patients.
Conclusions:
- The new phantom is suitable for Quality Assurance (QA) of Cobalt-60 HDR and Iridium-192 HDR systems.
- It enhances the confirmation of accurate dose distribution in GYN implants, including non-traditional ones.