SU-E-T-265: Reducing Pacemaker Doses with a Lead Sheet: A Multi-Detector Study
A Bourgouin1,2,3, N Varfalvy1,2,3, L Beaulieu1,2,3
1CHUQ Pavillon Hotel-Dieu de Quebec, Québec, Québec.
Lead shielding can effectively reduce radiation dose to pacemakers during antero-posterior radiotherapy. However, its use is not advised for postero-anterior fields, and placement outside the radiation field is crucial for optimal benefit.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Dosimetry
Background:
- Pacemakers are sensitive to radiation, necessitating dose reduction strategies during radiotherapy.
- Lead shielding is a potential method for mitigating radiation exposure to implanted medical devices.
Purpose of the Study:
- To evaluate the efficacy and safety of lead shielding for reducing radiation dose to pacemakers.
- To assess the impact of lead shielding placement and beam energy on dose reduction.
Main Methods:
- Measurements were conducted using a solid water phantom at 0.5 cm depth with 6 MV and 23 MV photon beams.
- Dosimetry was performed using a parallel plate ion chamber and a plastic scintillation detector (PSD) prototype.
- Lead shielding (1.6 mm) was tested with various field sizes and positions relative to the phantom.
Main Results:
- Both detectors showed excellent agreement (0.6%) across all measurements.
- Antero-posterior fields at 23 MV showed the greatest dose reduction (65%) with shielding outside the field.
- Shielding inside the field increased the dose for both antero-posterior and postero-anterior beams; benefits diminished beyond 35 cm from the field edge.
Conclusions:
- Lead shielding is recommended for antero-posterior radiotherapy to decrease pacemaker dose.
- For postero-anterior fields, lead shielding should be avoided or placed outside the treatment field.
- Plastic scintillation detectors (PSD) are suitable for in vivo monitoring of radiation dose to pacemakers and other sensitive sites like fetuses.