SU-E-T-236: Brachytherapy for Breast Cancer Patients with Implanted Pacemeker, Using Multi Lumen Partial Breast
1Christiana Care Hospital, Newark, DE.
Accelerated partial breast irradiation with a SAVI applicator safely treats breast cancer in patients with pacemakers. Advanced planning tools ensured the pacemaker received a dose below the 2Gy limit, maintaining normal function.
Area of Science:
- Oncology
- Medical Physics
- Radiotherapy
Background:
- Implanted pacemakers pose a challenge for breast cancer radiotherapy, particularly Accelerated Partial Breast Irradiation (APBI).
- High-dose-rate brachytherapy requires careful planning to avoid overdosing sensitive implanted devices.
Purpose of the Study:
- To discuss the optimized use of a Strut-Adjusted Volume Implant (SAVI) applicator for APBI in patients with pacemakers.
- To ensure effective tumor coverage without exceeding the dose limit for the pacemaker and its leads.
Main Methods:
- A patient with a right-sided pacemaker underwent breast-conserving surgery and High Dose Rate (HDR) brachytherapy using an Ir-192 source.
- A seven-catheter SAVI device was implanted, and a 3D treatment plan was generated using the Oncentra Brachy system.
- Optimization tools, including Inverse Planning Simulated Annealing (IPSA), were used to minimize the dose to the pacemaker.
Main Results:
- The SAVI device allowed for the delivery of APBI while keeping the pacemaker/lead dose below the 2Gy threshold.
- Achieved target coverage: V100=90.1%, V95=95.5%, V90=98.5%.
- The average dose to the pacemaker assembly over ten fractions was 140cGy, with no observed change in pacemaker function.
Conclusions:
- Optimized HDR brachytherapy with a SAVI applicator is a safe and effective treatment option for breast cancer patients with implanted pacemakers.
- Advanced planning and optimization tools enable the delivery of radiation without compromising pacemaker functionality.