SU-E-T-223: High-Energy Photon Standard Dosimetry Data: A Quality Assurance Tool
J Lowenstein1, S Kry1, A Molineu1
1UT MD Anderson Cancer Center, Houston, TX.
Purpose:
Describe the Radiological Physics Center's (RPC) extensive standard dosimetry data set determined from on-site audits measurements.
Method And Materials:
Measurements were made during on-site audits to institutions participating in NCI funded cooperative clinical trials for 44 years using a 0.6cc cylindrical ionization chamber placed within the RPC's water tank. Measurements were made on Varian, Siemens, and Elekta/Philips accelerators for 11 different energies from 68 models of accelerators. We have measured percent depth dose, output factors, and off-axis factors for 123 different accelerator model/energy combinations for which we have 5 or more sets of measurements. The RPC analyzed these data and determined the 'standard data' for each model/energy combination. The RPC defines 'standard data' as the mean value of 5 or more sets of dosimetry data or agreement with published depth dose data (within 2%).
Results:
The analysis of these standard data indicates that for modern accelerator models, the dosimetry data for a particular model/energy are within ï,±2%. The RPC has always found accelerators of the same make/model/energy combination have the same dosimetric properties in terms of depth dose, field size dependence and off-axis factors. Because of this consistency, the RPC can assign standard data for percent depth dose, average output factors and off-axis factors for a given combination of energy and accelerator make and model.
Conclusions:
The RPC standard data can be used as a redundant quality assurance tool to assist Medical Physicists to have confidence in their clinical data to within 2%. The next step is for the RPC to provide a way for institutions to submit data to the RPC to determine if their data agrees with the standard data as a redundant check. This work was supported by PHS grants CA10953 awarded by NCI, DHHS.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Atomic Emission Spectroscopy: Overview
Biological Effects of Radiation
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Dose Size and Dosing Frequency: Determination Methods