Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Therapists' Attitudes and Exclusion Criteria for Prolonged Exposure and EMDR in Patients With Eating Disorders and PTSD.

European eating disorders review : the journal of the Eating Disorders Association·2025
Same author

French survey found that the use of clonidine had increased in neonatal intensive care units.

Acta paediatrica (Oslo, Norway : 1992)·2024
Same author

Clinical Outcomes of 5000 IU Heparin Versus Activated Clotting Time-Guided Heparinization During Noncardiac Arterial Procedures: A Propensity Score Matched Analysis.

Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists·2024
Same author

Decision-making for extremely preterm infants with severe hemorrhages on head ultrasound: Science, values, and communication skills.

Seminars in fetal & neonatal medicine·2023
Same author

Effects of iron supplementation on microcytic and hypochromic red blood cells during the third trimester of pregnancy.

International journal of laboratory hematology·2022
Same author

2017 French health authority guidelines for neonates at risk of early-onset sepsis: Adherence and barriers in a tertiary hospital.

Archives de pediatrie : organe officiel de la Societe francaise de pediatrie·2022

Related Experiment Video

Updated: Mar 2, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
10:17

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry

Published on: April 23, 2019

10.4K

Multicenter verification of the Sysmex XN-Series.

M Schoorl1, M Schoorl1, M Chevallier1

  • 1Department of Clinical Chemistry, Haematology & Immunology, Northwest Clinics, Alkmaar, the Netherlands.

International Journal of Laboratory Hematology
|May 19, 2017
PubMed
Summary

Verification of Sysmex XN hematology modules showed uniform results across multiple labs. This straightforward method confirmed consistency with the previous Sysmex XE-2100 analyzer, ensuring reliable patient data.

Keywords:
Sysmex XNhematology analyzerperformanceverification

More Related Videos

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
05:58

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes

Published on: March 22, 2022

4.6K
Multimodality Diagnosis of Mesenteric Ischemia
05:07

Multimodality Diagnosis of Mesenteric Ischemia

Published on: July 21, 2023

1.3K

Related Experiment Videos

Last Updated: Mar 2, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
10:17

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry

Published on: April 23, 2019

10.4K
Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
05:58

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes

Published on: March 22, 2022

4.6K
Multimodality Diagnosis of Mesenteric Ischemia
05:07

Multimodality Diagnosis of Mesenteric Ischemia

Published on: July 21, 2023

1.3K

Area of Science:

  • Hematology
  • Laboratory Medicine
  • Medical Device Verification

Background:

  • Hemocytometry equipment verification is challenging due to lack of control materials and need for fresh samples.
  • Harmonization and uniformity are critical, especially in multicenter settings and across laboratory departments.
  • Current verification often relies on inter-equipment comparison and reproducibility assessments.

Purpose of the Study:

  • To evaluate the performance and uniformity of seven Sysmex XN hematology modules.
  • To establish a straightforward verification method for multicenter laboratory harmonization.
  • To ensure consistency of results with a previously validated analyzer (Sysmex XE-2100).

Main Methods:

  • Seven Sysmex XN hematology modules across three laboratories were tested with 160 samples.
  • Results were initially compared to a Sysmex XE-2100 analyzer using linear regression.
  • Bland-Altman analysis with adjusted limits of agreement was employed for detailed evaluation.
  • Reproducibility was assessed 10 times for each parameter using patient samples at varying levels.

Main Results:

  • Linear regression showed excellent correlation, but slope/intercept criteria were often unmet.
  • Bland-Altman analysis demonstrated that over 90% of determinations met the agreement criteria for most parameters.
  • Only Red Blood Cell (RBC) and Hematocrit (Ht) required recalibration in a few modules.
  • All tested parameters exhibited reproducibility within manufacturer specifications and biological variability.

Conclusions:

  • The evaluated Sysmex XN hematology modules demonstrated uniform and consistent performance.
  • The straightforward verification method confirmed results identical to the established Sysmex XE-2100 analyzer.
  • This approach facilitates harmonization and ensures reliable diagnostic data in multicenter laboratory settings.