Multicenter verification of the Sysmex XN-Series
M Schoorl1, M Schoorl1, M Chevallier1
1Department of Clinical Chemistry, Haematology & Immunology, Northwest Clinics, Alkmaar, the Netherlands.
International Journal of Laboratory Hematology
|May 19, 2017
Summary
Verification of Sysmex XN hematology modules showed uniform results across multiple labs. This straightforward method confirmed consistency with the previous Sysmex XE-2100 analyzer, ensuring reliable patient data.
Area of Science:
- Hematology
- Laboratory Medicine
- Medical Device Verification
Background:
- Hemocytometry equipment verification is challenging due to lack of control materials and need for fresh samples.
- Harmonization and uniformity are critical, especially in multicenter settings and across laboratory departments.
- Current verification often relies on inter-equipment comparison and reproducibility assessments.
Purpose of the Study:
- To evaluate the performance and uniformity of seven Sysmex XN hematology modules.
- To establish a straightforward verification method for multicenter laboratory harmonization.
- To ensure consistency of results with a previously validated analyzer (Sysmex XE-2100).
Main Methods:
- Seven Sysmex XN hematology modules across three laboratories were tested with 160 samples.
- Results were initially compared to a Sysmex XE-2100 analyzer using linear regression.
- Bland-Altman analysis with adjusted limits of agreement was employed for detailed evaluation.
- Reproducibility was assessed 10 times for each parameter using patient samples at varying levels.
Main Results:
- Linear regression showed excellent correlation, but slope/intercept criteria were often unmet.
- Bland-Altman analysis demonstrated that over 90% of determinations met the agreement criteria for most parameters.
- Only Red Blood Cell (RBC) and Hematocrit (Ht) required recalibration in a few modules.
- All tested parameters exhibited reproducibility within manufacturer specifications and biological variability.
Conclusions:
- The evaluated Sysmex XN hematology modules demonstrated uniform and consistent performance.
- The straightforward verification method confirmed results identical to the established Sysmex XE-2100 analyzer.
- This approach facilitates harmonization and ensures reliable diagnostic data in multicenter laboratory settings.