SU-D-217A-03: Nuclear Medicine Uniformity Assessment Using 2D Noise Power Spectrum
J Nelson1,1,2,3,4,5,6,1, O Christianson1,1,2,3,4,5,6,1, B Harkness1,1,2,3,4,5,6,1
1Duke University Medical Center, Durham, NC.
Purpose:
Nuclear medicine quality control programs require daily evaluation for the presence of potential non-uniformities by commonly utilizing a traditional pixel value-based assessment (Integral CFOVUniformity). While this method effectively captures regional non- uniformities in the image, it does not adequately reflect subtle periodic structures that are visually apparent and clinically unacceptable, therefore requiring the need for additional visual inspection of the image. The goal of this project was to develop a new uniformity assessment metric by targetingstructural patterns and more closely correlating with visual inspection.
Methods:
The new quantitative uniformity assessment metric is based on the 2D Noise Power Spectrum (NPS). A full 2D NPS was performed on each image. The NPS was thresholded to remove quantum noise and further filtered by the visual response function. A score, the Structure Noise Index (SNI), was then applied to each based on the average magnitude of the structured noise in the processed image. To verify the validity of the new metric, 50 daily uniformity images with varying degrees of visual structured and non-structured non-uniformity were scored by 5 expert nuclear medicine physicists. The correlation between the visual score and SNI were assessed. The Integral CFOV was also compared against the visual score.
Results:
Our new SNI assessment metric compared to the Integral CFOV showed in increase in sensitivity from 67% to 100% in correctly identifying structured non-uniformities. The overall positive predictive value also increased from 55% to 72%.
Conclusions:
Our new uniformity metric correlates much more closely with visual assessment of structured non- uniform NM images than the traditional pixel-based method. Using this new metric in conjunction with the traditional pixel value-based assessment will allow a more accurate quantitative assessment of nuclear medicineuniformity.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Two-Dimensional (2D) NMR: Overview
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse....
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...
2D NMR: Heteronuclear Single-Quantum Correlation Spectroscopy (HSQC)
Nuclear Overhauser Enhancement (NOE)